Publication: Meniscal repair using large diameter horizontal sutures increases fixation strength: an in vitro study
Loading...
Date
Advisor
Journal Title
Journal ISSN
Volume Title
Publisher
Wiley
Type
Abstract
AbstractPurposeThe purpose of this study was to compare the mechanical characteristics of meniscal repair fixation using horizontal sutures and two different diameter sutures under submaximal cyclic and load to failure test conditions.MethodA 2‐cm long anteroposterior vertical longitudinal incision was created in two groups of bovine medial menisci. Lesions were repaired using either #2‐0 (Group 1), or #2 (Group 2) Fiberwire suture. Following repair, the lesion was extended through the posterior and anterior meniscal horns so that no tissue secured the repair site. Specimens underwent submaximal cyclic (5–50 N at 1 Hz for 500 cycles) and load to failure testing (5 mm/min crosshead speed) in a servo hydraulic device. Specimen failure mode was verified by the primary investigator. An alpha level of P < 0.05 was selected to indicate statistical significance.ResultsGroup 2 displayed greater load at failure (132.1 ± 54.4 N) than Group 1 (91.9 ± 26.2 N) (P = 0.02). Group 2 also displayed greater stiffness (47.1 ± 8.3 N/mm) than Group 1 (38.5 ± 10.2 N/mm) (P = 0.03). The failure mode for all specimens was suture pull‐through the meniscal tissue. Larger diameter suture provided superior mechanical meniscal fixation.ConclusionIf Horizontal suture would be used in meniscal repair, the most suitable larger diameter suture should be used.
Description
Subject
Equipment Failure Analysis, Sutures, Materials Testing, Animals, Cattle, Equipment Design, Menisci, Tibial, Tibial Meniscus Injuries