Sensitivity and sensing depth analysis of open-ended contact probes for cancer diagnosis

dc.contributor.author Aydınalp, Cemanur
dc.contributor.author joof, Sulayman
dc.contributor.author Akduman, İbrahim
dc.contributor.author Yılmaz, Tuba
dc.contributor.department Elektronik ve Haberleşme Mühendisliği tr_TR
dc.contributor.department Electronics and Communication Engineering tr_TR
dc.date.accessioned 2019-10-07T12:12:36Z
dc.date.available 2019-10-07T12:12:36Z
dc.date.issued 2019
dc.description.abstract Open-ended contact probes have been widely utilized in laboratory environment to quantify the dielectric properties of high permittivity and lossy materials such as biological tissues and other lossy liquids. At microwave frequencies, the dielectric properties of biologi- cal tissues have been quantified with open-ended contact probes method with the motivation of building microwave diagnostic and therapeutic technologies. The method is preferred to other microwave dielectric property measurement techniques due to broadband measurement capabil- ities and minimal sample preparation requirements. However, the measurement procedure is cumbersome and the measurements tend to be error-prone. The commercial probes can have error rates as high as 5% and it can increase with the cable movements, calibration degradation, and probe wear-off. Although the technique is powerful and able to measure the inherent dielec- tric property discrepancy between different tissues the applications are confined to laboratory use. To explore the true potential of this technique and enable the transition of this technique to a diagnostic technology, there is a need to improve the technique and analyze the performance in a layered heterogeneous medium. In this work, we performed simulations and measurements with open-ended contact probes having different aperture diameters. The probes are built with copper and Teflon sandwiched between inner and outer conductors. Measurements with open- ended coaxial probes are performed on layered medium emulating the dielectric properties of the biological tissues to analyze the sensing depth and sensitivity of four in-house fabricated open- ended coaxial probes. An in-house algorithm is utilized to retrieve the dielectric properties of the mediums utilized during this study. tr_TR
dc.identifier.citation Cemanur Aydinalp, Sulayman Joof, Ibrahim Akduman, and Tuba Yilmaz, " Sensitivity and Sensing Depth Analysis of Open-ended Contact Probes for Cancer Diagnosis." 2019 Progress in Electromagnetics Research Symposium (PIERS-Rome). IEEE, 2019. tr_TR
dc.identifier.uri http://hdl.handle.net/11527/18063
dc.language.iso en tr_TR
dc.publisher İstanbul Teknik Üniversitesi tr_TR
dc.subject Cancer diagnosis en_US
dc.title Sensitivity and sensing depth analysis of open-ended contact probes for cancer diagnosis en_US
dc.type Conference Paper en_US
Dosyalar
Orijinal seri
Şimdi gösteriliyor 1 - 1 / 1
thumbnail.default.alt
Ad:
version2_amended.pdf
Boyut:
473.72 KB
Format:
Adobe Portable Document Format
Açıklama
Konferans bildirisi
Lisanslı seri
Şimdi gösteriliyor 1 - 1 / 1
thumbnail.default.placeholder
Ad:
license.txt
Boyut:
3.06 KB
Format:
Item-specific license agreed upon to submission
Açıklama