Yayın:
Loop security and tensile properties of polyblend and traditional suture materials

Yükleniyor...
Küçük Resim

Kurum Yazarları

Item type:Araştırmacı/Yazar,
Bozdağ, Süreyya Ergün
Öğretim Görevlisi Dr.
Item type:Araştırmacı/Yazar,
Sünbüloğlu, Emin
Profesor

Danışman

Bölüm / Program

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Wiley

Türü

Araştırma Projeleri

Akademik Birimler

Dergi Sayısı

Özet

AbstractTensile and knot properties of new generation (polyblend) and traditional suture materials in orthopedic surgery were investigated in standard laboratory conditions. Study focused on Fiberwire No. 5 and 2, Ethibond No. 5, 2 and 00, Orthocord No. 2, MaxBraid No. 2, Prolene No. 0 and 00, PDS No. 0 and 00, and Vicryl No. 2, 0 and 00. A 27‐cm suture loop was fastened with 10 knots for ten samples for each type. Test parameters were tensile load to failure, elongation at failure point and knot slippage, and volume of 10‐fold knots. Results were compared using ANOVA test. Failure load of No. 5 Fiberwire (625.0 ± 30.0 N) was significantly higher compared to all other suture types. Tensile strengths of MaxBraid No. 2 (287 ± 11 N) was significantly stronger compared to two other No. 2 polyblend sutures types and Ethibond No. 5. Knot slippage of Fiberwire No. 5 (14 ± 1.9 mm) was significantly higher compared to all other suture types. Ethibond No. 2 (0.1 ± 0.3 mm) had the lowest knot slippage. Elongation at the failure point of Fiberwire No. 2 (5%) was significantly lower than all other suture types. Mean calculated knot volume of #5 Fiberwire (73 ± 6.9 mm3) was significantly higher compared to #5 Ethibond (53 ± 4.8 mm3). Results of the study proved presence of significant differences between tensile and knot properties of various suture types and sizes. Loop security of larger diameter sutures is not always higher than thinner sutures. Suture elongation and knot slippage are important failure modes for high‐diameter sutures and short‐suture loops.

Tanım

Dergi veya Seri

Knee Surgery, Sports Traumatology, Arthroscopy

ISSN

0942-2056

ISBN

Haklar

CLOSED

Anahtar Kelimeler

Sutures, Polyethylene Terephthalates, Loop security, Suture Techniques, Knot security, Soft tissue, Polypropylenes, Tensile Strength, Materials Testing, Surgery, Suture material, Polyglactin 910, Tendon, Repair

Alıntı

Koleksiyonlar

Onay

Gözden geçir

Tamamlayıcı Bilgiler

Referans Gösteren

Related Patent

Related Goal

4
Görüntülenme
0
İndirme
Altmetric
Dimensions
PlumX Metrikleri
BIP! Indicators
Google Scholar
Scholar'da Ara ↗