Yayın:
A novel preconcentration method for determination of iron and lead using Chromosorb-103 and flame atomic absorption spectrometry

dc.contributor.authorBakircioglu, Y.
dc.contributor.authorBakircioglu, D.
dc.contributor.authorTokman, N.
dc.date.accessioned2026-01-26T06:53:28Z
dc.date.issued2005-08-01
dc.description.abstractSyringe connected-minicolumn (SCM) method for the preconcentration/separation of iron and lead prior to their determination by flame atomic absorption spectrometry (FAAS) was developed. The proposed method is substituted for classical batch and column methods. The method proposed was compared with the column method with respect to easiness, rapidness, simplicity and some analytical performance criteria such as recovery, precision, accuracy and risk of contamination. A minicolumn was filled with Chromosorb-103 as a sorbent and connected with a syringe. The experimental conditions such as pH of sample, concentration and amount of eluent, flow rates of sample and eluent, concentration of 8-hydroxyquinoline (oxine) as a complexing agent were optimized. The sample solution with or without oxine was drawn into the syringe and discharged manually passing through the resin. Analytes were retained at pH ≥5 and eluted with 1 M nitric acid in acetone quantitatively. Analyte elements could be concentrated up to 200-fold. The iron and lead in river-water samples and in certified reference Bovine Liver (NIST 1577b) were quantitatively recovered with relative standard deviation lower than 10%.
dc.description.urihttps://doi.org/10.1016/j.aca.2005.01.056
dc.description.urihttps://dx.doi.org/10.1016/j.aca.2005.01.056
dc.description.urihttps://hdl.handle.net/20.500.14551/24126
dc.identifier.doi10.1016/j.aca.2005.01.056
dc.identifier.endpage30
dc.identifier.issn0003-2670
dc.identifier.openairedoi_dedup___::f2979a21a915661e64b456ea092341c4
dc.identifier.orcid0000-0003-3013-5793
dc.identifier.orcid0000-0003-3384-0340
dc.identifier.startpage26
dc.identifier.urihttps://hdl.handle.net/11527/63213
dc.identifier.volume547
dc.language.isoeng
dc.publisherElsevier BV
dc.relation.ispartofAnalytica Chimica Acta
dc.rightsCLOSED
dc.sdg.typeGoal 6: Clean Water and Sanitation
dc.subjectBovine Liver
dc.subjectFlame Atomic Absorption Spectrometry
dc.subjectSolid Phase Extraction
dc.subjectTrace-Elements
dc.subjectWater
dc.subjectChromosorb-103
dc.subjectAqueous Samples
dc.subjectSyringe-Connected Minicolumn Method
dc.subjectSyringe
dc.subjectNickel
dc.subjectFlow-Injection
dc.subjectSolid-Phase Extraction
dc.subjectChromatographic Stationary Phases
dc.subjectSilica-Gel
dc.subjectCadmium
dc.titleA novel preconcentration method for determination of iron and lead using Chromosorb-103 and flame atomic absorption spectrometry
dc.typeArticle
dspace.entity.typePublication

Dosyalar

Koleksiyonlar