The use of solid-phase extraction and direct injection of a copolymer sorbent as slurry into the graphite furnace prior to determination of cadmium by ETAAS

Loading...
Thumbnail Image

Institution Authors

Advisor

Department

Journal Title

Journal ISSN

Volume Title

Publisher

Informa UK Limited

Research Projects

Organizational Units

Journal Issue

Abstract

Cadmium in different certified materials was determined by direct injection of sorbent loaded with the analyte into the graphite furnace of electrothermal atomic absorption spectrophotometer. For this purpose, cadmium was collected on ethylene glycol dimethacrylatemethacrylic acid copolymer treated with ammonium pyrolidine dithiocarbamate by the batch technique. After separation of liquid phase, slurry of the sorbent was prepared for the direct injection of the sorbent into graphite furnace. Optimum conditions for quantitative sorption and those for preparing homogeneous and durable slurry were investigated. A 100-fold enrichment factor could be reached. Cadmium in certified sea water and bovine liver samples was determined within the 95% confidence limits. The proposed technique is fast and simple, and the risk of contamination is low. The calculated detection limit for cadmium using sample-matching blanks was 0.09 µg L−1 according to the 3σ concept for the described method.

Description

Journal or Series

International Journal of Environmental Analytical Chemistry

ISSN

0306-7319

ISBN

Rights

Keywords

Citation

Collections

Endorsement

Review

Supplemented By

Referenced By

Related Patent

Related Goal

1
Görüntülenme
0
İndirme
Altmetric
Dimensions
PlumX Metrikleri
BIP! Indicators
Google Scholar
Scholar'da Ara ↗