Publication:
Determination of lead in rice grains by solid sampling HR-CS GFAAS

Loading...
Thumbnail Image

Advisor

Journal Title

Journal ISSN

Volume Title

Publisher

Elsevier BV

Research Projects

Organizational Units

Journal Issue

Abstract

A study was performed for the determination of lead in rice grains directly by solid sampling high resolution continuum source graphite furnace atomic absorption spectrometry (HR-CS GFAAS) without digesting sample. The effects of sample amount, pyrolysis/atomization temperatures on the determination of lead in rice were investigated and optimized using a certificated rice flour. The lead concentrations were determined applying 800 °C of pyrolysis and 1800 °C of atomization temperatures without modifier. LOD (N=10, 3σ) and characteristic mass were 2.3 μg kg(-1) and 8.1 pg, respectively. The certified lead value of a rice flour CRM was found in its uncertainity limits. The lead contents of various rice samples obtained from markets in Turkey were between 0.009 and 0.162 mg kg(-1) which are in acceptable range. The average lead concentrations in various rice grains taken from the same package were significantly different from each other.

Description

Subject

Lead, Spectrophotometry, Atomic, Seeds, Food Contamination, Oryza

Citation