Publication:
Determination of Chlorine in Plastics via SrCl by Solid-Sampling High-Resolution Continuum Source Graphite Furnace Molecular Absorption Spectrometry

Loading...
Thumbnail Image

Institution Authors

Item type:Person,
Özbek, Nil
Docent

Advisor

Department

Journal Title

Journal ISSN

Volume Title

Publisher

Informa UK Limited

Research Projects

Organizational Units

Journal Issue

Abstract

AbstractSolid sampling high resolution continuum source molecule absorption spectrometry (SS-HR-CS MAS) was applied for the determination of chlorine in plastic using the strontium monochloride (SrCl) molecule. For this purpose, 10 µL of 20 g L−1 strontium (prepared from Sr(NO3)2) solution were pipetted with aqueous Cl standards or 0.05 to 4 mg of slivered plastic samples on a platform and introduced into the furnace. Chlorine was determined with the molecular absorption of SrCl at 635.862 nm using 1100 °C and 2200 °C for the pyrolysis and vaporization temperatures, respectively. Aqueous standards were used for calibration. The accuracy of the method was evaluated using a certified polyethylene reference material. The limit of detection and characteristic mass values of the method were 2.5 ng and 0.4 ng, respectively. The chlorine concentrations in various polyethylene beverage containers were determined.